356 Micro-Ion® Plus Vacuum Pressure Transducers
Model: 356-micro-ion-plus-modules
356 Micro-Ion® Plus Vacuum Pressure Transducers
Model: 356-micro-ion-plus-modules
Technical Specs
Sensor Type
Micro-Ion® Gauge
Measurement Range
1 x 10-9 Torr to atmosphere
1 x 10-9 mbar to atmosphere
1 x 10-7 Pa to atmosphere
(Air and N2)
Repeatability
<5% of reading from 1 x 10-8 Torr to 50 Torr
Accuracy
<15% of reading from 1 x 10-8 Torr to 50 Torr (N2)
Filament
Yttria-coated iridium
Operating Temperature
10 - 40°C
Bake Out Temperature
150°C (with electronics removed)
Storage Temperature
-40 to 70°C
Relays
Two single-pole, double-throw relays (SPDT)
Relay Contact Rating
1 A at 30 VDC resistive, non-inductive
Relay Response
1 x 10-9 to 50 Torr; 1 x 10-9 to 70 mbar; 1 x 10-7 Pa to 7 kPa
Power Requirements
24 VDC ±15%, 26 W max
Exposed Materials
304 stainless steel, tantalum, tungsten, nickel, nickel iron alloy, borosilicate glass
Case Material
Aluminum extrusion
Internal Volume
10.8 cc
Weight
567 gm (20 oz) with NW16KF flange
Display
2 digits plus exponent, green LED (optional)
Ion Gauge Degas
Electron bombardment, 3.75 W with 2-minute timer
Ion Gauge Emission Current
Autoranging
Ion Gauge Auto On Setting
2 x 10-2 Torr; 3 x 10-2 mbar; 3 Pa, on decreasing pressure
Ion Gauge Auto Off Setting
3 x 10-2 Torr; 4 x 10-2 mbar; 4 Pa, on increasing pressure
Switch to Low Emission Setting
1 x 10-5 Torr; 1 x 10-5 mbar; 1 x 10-3 Pa, on increasing pressure
X-ray Limit
<3 x 10-10 Torr; <4 x 10-10 mbar; <4 x10-8 Pa
Compliance
CE
Technical Specs
Sensor Type
Micro-Ion® Gauge
Measurement Range
1 x 10-9 Torr to atmosphere
1 x 10-9 mbar to atmosphere
1 x 10-7 Pa to atmosphere
(Air and N2)
Repeatability
<5% of reading from 1 x 10-8 Torr to 50 Torr
Accuracy
<15% of reading from 1 x 10-8 Torr to 50 Torr (N2)
Filament
Yttria-coated iridium
Operating Temperature
10 - 40°C
Bake Out Temperature
150°C (with electronics removed)
Storage Temperature
-40 to 70°C
Relays
Two single-pole, double-throw relays (SPDT)
Relay Contact Rating
1 A at 30 VDC resistive, non-inductive
Relay Response
1 x 10-9 to 50 Torr; 1 x 10-9 to 70 mbar; 1 x 10-7 Pa to 7 kPa
Power Requirements
24 VDC ±15%, 26 W max
Exposed Materials
304 stainless steel, tantalum, tungsten, nickel, nickel iron alloy, borosilicate glass
Case Material
Aluminum extrusion
Internal Volume
10.8 cc
Weight
567 gm (20 oz) with NW16KF flange
Display
2 digits plus exponent, green LED (optional)
Ion Gauge Degas
Electron bombardment, 3.75 W with 2-minute timer
Ion Gauge Emission Current
Autoranging
Ion Gauge Auto On Setting
2 x 10-2 Torr; 3 x 10-2 mbar; 3 Pa, on decreasing pressure
Ion Gauge Auto Off Setting
3 x 10-2 Torr; 4 x 10-2 mbar; 4 Pa, on increasing pressure
Switch to Low Emission Setting
1 x 10-5 Torr; 1 x 10-5 mbar; 1 x 10-3 Pa, on increasing pressure
X-ray Limit
<3 x 10-10 Torr; <4 x 10-10 mbar; <4 x10-8 Pa
Compliance
CE
Resources & Downloads
Device Description Files
Other