226B Baratron® 0.2-1000 Torr Differential Capacitance Manometers
Model: 226B21MGAGAFB1T1
226B Baratron® 0.2-1000 Torr Differential Capacitance Manometers
Model: 226B21MGAGAFB1T1

Overview
Overview
The 226B21MGAGAFB1T1 Baratron® differential capacitance manometer is ranged for 20 mbar, has ISO-KF NW16 reference and measurement side fittings, 0.50% of Full Scale accuracy, ±15 VDC input, 0 - 1 VDC analog output with bidirectional calibration, a 5-post berminal block adaptor electrical connector, and a standard mounting bracket. It is designed for applications in industry, electronics manufacturing, and analytical equipment. Its rugged, industrial-grade design is suitable for use in applications with high levels of RF/EM interference.
Technical Specs
- TypeVacuum/pressure Transducer
- Measurement TypeDifferential
- Full Scale Range20 mbar
- Analog Output0-1 VDC
- CalibrationBidirectional
- Electrical Connector5-pin Terminal block adapter
- Resolution0.01% of Full Scale
- Accuracy0.50% of Full Scale
- Zero Temperature Coefficient0.1% Full Scale/°C for standard accuracy specification
- Span Temperature Coefficient0.04% of Reading/°C
- Sensor Internal VolumeMeasurement Side: 1.4 cm3, Reference Side: 9.0 cm3
- Operating Temperature Range0-50°C
- Maximum OverpressureMeasurement Side: 120% of Full Scale or 20 psi (140 kPa), whichever is higher, Reference Side: 120% of Full Scale
- Maximum Line Pressure40 psig (275 kPa)
- Exposed MaterialsMeasurement Side: Inconel, Reference Side: Inconel, ceramic, palladium, stainless steel, glass
- Measurement Side FittingNW16-KF
- Reference Side FittingNW16-KF
- Mounting BracketStandard
- Trip Points1
- Power Requirements±15 VDC
- ComplianceCE, SEMI S2-0706
Technical Specs
- TypeVacuum/pressure Transducer
- Measurement TypeDifferential
- Full Scale Range20 mbar
- Analog Output0-1 VDC
- CalibrationBidirectional
- Electrical Connector5-pin Terminal block adapter
- Resolution0.01% of Full Scale
- Accuracy0.50% of Full Scale
- Zero Temperature Coefficient0.1% Full Scale/°C for standard accuracy specification
- Span Temperature Coefficient0.04% of Reading/°C
- Sensor Internal VolumeMeasurement Side: 1.4 cm3, Reference Side: 9.0 cm3
- Operating Temperature Range0-50°C
- Maximum OverpressureMeasurement Side: 120% of Full Scale or 20 psi (140 kPa), whichever is higher, Reference Side: 120% of Full Scale
- Maximum Line Pressure40 psig (275 kPa)
- Exposed MaterialsMeasurement Side: Inconel, Reference Side: Inconel, ceramic, palladium, stainless steel, glass
- Measurement Side FittingNW16-KF
- Reference Side FittingNW16-KF
- Mounting BracketStandard
- Trip Points1
- Power Requirements±15 VDC
- ComplianceCE, SEMI S2-0706