228A Baratron® 0.2-1000 Torr Differential Capacitance Manometers with 2 Trip Points
Model: 228A12KBGBGFB7C1BB
228A Baratron® 0.2-1000 Torr Differential Capacitance Manometers with 2 Trip Points
Model: 228A12KBGBGFB7C1BB

Overview
Overview
The 228A12KBGBGFB7C1BB Baratron® differential capacitance manometer has a 100 kPa full scale pressure range with a 0-5 VDC analog output with bidirectional calibration. It features 3/16 in. OD tube fittings on both sides. It provides two (2) independently-adjustable trip relays for control of external equipment and components. The patented capacitance sensor is built entirely from Inconel® nickel alloys on its measurement side, which offers superior corrosion resistance over long periods of time. Because the sensor operates by measuring the capacitance shift between a diaphragm exposed to the process and an electrode disk (rather than measuring the property of the gas), it is not sensitive to gas composition, and thus eliminates the need for gas-specific correction factors. The product can be used to measure either the true differential pressure or vacuum between two locations, or the reference side of the product can be left open to provide a true reference to local atmospheric pressure. Applications include air and gas flow measurements for filters and analytical systems, downstream pressure control in thin film processing systems, and automated leak testing systems.
Technical Specs
- TypePressure Transducer
- Sensor TypeCapacitance Manometer
- Measurement TypeDifferential
- Full Scale Range100 kPa
- Analog Output0 - 5 VDC
- CalibrationBidirectional
- Fitting Type0.1875 in. OD tube
- Electrical Connector15-pin HD D-sub
- Resolution0.01% of Full Scale
- Accuracy0.50% of Full Scale
- Zero Temperature Coefficient0.1% Full Scale/°C for standard accuracy specification
- Span Temperature Coefficient0.04% of Reading/°C
- Sensor Internal VolumeMeasurement Side" 1.4 cm3, Reference Side 9.0 cm3
- Operating Temperature Range0 - 50°C
- Maximum OverpressureMeasurement Side: 120% of Full Scale or 20 psi (140 kPa), whichever is higher, Reference Side: 120% of Full Scale
- Maximum Line Pressure40 psig (275 kPa)
- Exposed MaterialsMeasurement Side: Inconel, Reference Side: Inconel, ceramic, palladium, stainless steel, glass
- Measurement Side Fitting0.1875 in. OD tube
- Reference Side Fitting0.1875 in. OD tube
- Mounting BracketStandard
- Trip Points2
- Trip RelaysIndependently adjustable from -100% to +100% of Full Scale
- Relay Contact RatingsDPDT, 1.0 A at 30 VDC or 0.3 amps at 120 VAC
- Warm-up Time30 minutes
- Power Requirements24 VDC
- ComplianceCE, SEMI S2-0706
Technical Specs
- TypePressure Transducer
- Sensor TypeCapacitance Manometer
- Measurement TypeDifferential
- Full Scale Range100 kPa
- Analog Output0 - 5 VDC
- CalibrationBidirectional
- Fitting Type0.1875 in. OD tube
- Electrical Connector15-pin HD D-sub
- Resolution0.01% of Full Scale
- Accuracy0.50% of Full Scale
- Zero Temperature Coefficient0.1% Full Scale/°C for standard accuracy specification
- Span Temperature Coefficient0.04% of Reading/°C
- Sensor Internal VolumeMeasurement Side" 1.4 cm3, Reference Side 9.0 cm3
- Operating Temperature Range0 - 50°C
- Maximum OverpressureMeasurement Side: 120% of Full Scale or 20 psi (140 kPa), whichever is higher, Reference Side: 120% of Full Scale
- Maximum Line Pressure40 psig (275 kPa)
- Exposed MaterialsMeasurement Side: Inconel, Reference Side: Inconel, ceramic, palladium, stainless steel, glass
- Measurement Side Fitting0.1875 in. OD tube
- Reference Side Fitting0.1875 in. OD tube
- Mounting BracketStandard
- Trip Points2
- Trip RelaysIndependently adjustable from -100% to +100% of Full Scale
- Relay Contact RatingsDPDT, 1.0 A at 30 VDC or 0.3 amps at 120 VAC
- Warm-up Time30 minutes
- Power Requirements24 VDC
- ComplianceCE, SEMI S2-0706
Resources & Downloads
Drawings & CADs
Literature
228A Baratron® Differential Capacitance Manometer Datasheet(239.6 kB, PDF)