475 Convectron® Vacuum Gauge Controller
Model: 475-convectron-vacuum-gauge-controller
475 Convectron® Vacuum Gauge Controller
Model: 475-convectron-vacuum-gauge-controller
Technical Specs
Type
Convectron® Gauge Controller
Measurement Range
1x10-4 Torr to Atmosphere
1x10-4 mbar to Atmosphere
1x10-2 Pa to Atmosphere
(N2)
Step Size at Minimum Pressure
1 x 10-4 Torr, 1 x 10-4 mbar, 1 x 10-2 Pa
Display
Vacuum Fluorescent
Display Update Rate
Every 0.5 sec
Units
mBar, Pascal, or Torr (optional)
Compatible with
Series 275 Convectron® Vacuum Gauges
Power Requirements
12 to 24 VDC, 6 W continuous
Communication Protocol
RS232 or RS485 (optional)
Relays
0 or 2 (optional)
Set Point Contact Rating
5 A @ 250 VAC resistive load
Set Point Configuration
Two single pole, double-throw (optional)
Set Point Range
1 x 10-3 to 1000 Torr
1 x 10-3 to 1333 mbar
1 x 10-1 Pa to 133 kPa
Set Point Resolution
2 significant digits
Operating Temperature
0° to 40°C
Storage Temperature
-40°C to 70°C
Weight
720 gm (25 oz)
Compliance
CE
Technical Specs
Type
Convectron® Gauge Controller
Measurement Range
1x10-4 Torr to Atmosphere
1x10-4 mbar to Atmosphere
1x10-2 Pa to Atmosphere
(N2)
Step Size at Minimum Pressure
1 x 10-4 Torr, 1 x 10-4 mbar, 1 x 10-2 Pa
Display
Vacuum Fluorescent
Display Update Rate
Every 0.5 sec
Units
mBar, Pascal, or Torr (optional)
Compatible with
Series 275 Convectron® Vacuum Gauges
Power Requirements
12 to 24 VDC, 6 W continuous
Communication Protocol
RS232 or RS485 (optional)
Relays
0 or 2 (optional)
Set Point Contact Rating
5 A @ 250 VAC resistive load
Set Point Configuration
Two single pole, double-throw (optional)
Set Point Range
1 x 10-3 to 1000 Torr
1 x 10-3 to 1333 mbar
1 x 10-1 Pa to 133 kPa
Set Point Resolution
2 significant digits
Operating Temperature
0° to 40°C
Storage Temperature
-40°C to 70°C
Weight
720 gm (25 oz)
Compliance
CE