The 631F11TBFP Heated Absolute Baratron® Capacitance Manometer features 10 Torr Full Scale pressure measurement, is heated to 200°C, and has a female 8 VCR type fitting. It operates at very high internal temperatures, minimizes the deposition of process byproducts within the sensor and nearly eliminating output drift caused by contamination. Its 200°C operating temperature permits use in the most demanding vacuum processes in semiconductor manufacturing such as metal etching and nitride film chemical vapor deposition (CVD). Its new packaging reduces the amount of installation space needed by at least 40%, making it even easier to design into new systems or retrofit into existing equipment. The 631F is completely self-contained, requiring only input power of ±15VDC for operation. No separate electronics modules are used, which both reduces the amount of installation space needed and provides better performance due to elimination of noise generating interconnecting cables. Its Inconel® sensor offers extremely high resistance to corrosion, and its overpressure rating of 45 psia ensures good repeatability and stability regardless of the system operating conditions. The Type 631F high-performance analog electronics also include fail-safe over temperature protection and LED status lamps to indicate its operating state.
Technical Specs
- TypeVacuum/pressure Transducer
- Measurement TypeAbsolute
- Full Scale Range10 Torr
- Sensor Temperature200°C
- Accuracy0.50% of Reading
- Fitting Type8 VCR female
- Analog Output0-10 VDC
- Analog Connector15-pin D-sub
- Trip Relays0
- Resolution0.01% of Full Scale
- Zero Temperature Coefficient0.008% of Full Scale/°C (0.016% of Full Scale/°C for 1 Torr range)
- Span Temperature Coefficient0.02% of Reading/°C
- Response Time50 msec
- Warm-up Time4 hours or less
- Internal Volume6.3 cm3
- Operating Temperature Range15-50°C
- Overpressure Limit45 psia
- Exposed MaterialsInconel® or Incoloy® nickel alloys
- Power Requirements± 15VDC (±5%) @ 1.0 amps max
- Ambient Airflow Requirements150 ft/min for ambient temperatures of 40 - 50°C
- ComplianceCE
Technical Specs
- TypeVacuum/pressure Transducer
- Measurement TypeAbsolute
- Full Scale Range10 Torr
- Sensor Temperature200°C
- Accuracy0.50% of Reading
- Fitting Type8 VCR female
- Analog Output0-10 VDC
- Analog Connector15-pin D-sub
- Trip Relays0
- Resolution0.01% of Full Scale
- Zero Temperature Coefficient0.008% of Full Scale/°C (0.016% of Full Scale/°C for 1 Torr range)
- Span Temperature Coefficient0.02% of Reading/°C
- Response Time50 msec
- Warm-up Time4 hours or less
- Internal Volume6.3 cm3
- Operating Temperature Range15-50°C
- Overpressure Limit45 psia
- Exposed MaterialsInconel® or Incoloy® nickel alloys
- Power Requirements± 15VDC (±5%) @ 1.0 amps max
- Ambient Airflow Requirements150 ft/min for ambient temperatures of 40 - 50°C
- ComplianceCE