Microvision 2 is the latest innovation in RGA technology from MKS, designed to meet all of the traditional requirements for an RGA sensor but with data collection at speeds unachievable with previous generation technologies. In addition, the Microvision 2 is designed to collect data at millisecond speeds per data point even when measuring data over the full dynamic range of the RGA. This capability has been achieved without sacrificing any of the necessary robustness, reliability and support which have made MKS the world leader in RGA products over the widest range of applications from semiconductor tools or particle accelerators to general industrial applications. Microvision 2 is useful in a wide range of situations where knowing the contents of your chamber is critical to the success of your process.
Technical Specs
- TypeResidual Gas Analyzer
- Measurement TechniqueMass Spectrometry
- Mass Range100, 200, or 300 amu
- DetectorFaraday
- Faraday1.5e-11 Torr (2e-11 mbar)
- FilamentsTungsten or Thoria coated Iridium
- SoftwareBuilt-in web applications allowing RGA control and data acquisition using a platform independent web browser without the need for dedicated installed software, EasyView, Process Eye Professional and TOOLweb RGA Windows software packages for Windows 2000, XP, or Vista
- Maximum Analyzer Operating Pressure7.6e-5 Torr (1e-4 mbar)
- Ion Source Sensitivity2e-4 A/mbar
- ResolutionBetter than 10% valley for peaks of equal height across mass range
- Bake Out Temperature250°C (with extender or electronics removed)
- Operating Temperature200°C Faraday only with electronics 10 - 40°C, 150°C Multiplier with electronics 10 - 40°C and pressure <1e-8 mbar, 90°C Multiplier with electronics 10 - 40°C and pressure <1e-5 mbar
- Cross Beam Ion SourceSingle, double, or triple, depending on configuration.
- UHV Ion SourceSingle, double, or triple, depending on configuration.
- UHV Out-gassing Rate<1e-9 mbar -l/s
- PVD Ion SourceSingle, double, or triple, depending on configuration.
- Zero Blast EffectDouble & Triple Filter: <5 ppm baseline levels next to mass 2 (not applicable with 5 in. single version)
- Maximum Operating Conditions10 - 40°C, 80% RH (non condensing)
- Microchannel plate3.8e-14 Torr (5e-14 mbar)
- Single channel EM7.6e-15 Torr (1e-14 mbar) triple configuration only
- Electronics Weight on Flange1.7 kg
- Power24 VDC, 3A external universal voltage supply (included)
- Data Acquisition MethodDedicated real time acquisition processor
- Data Acquisition Speed<3 ms per point for analog scanning
- Data Collection MethodsAnalog scanning, full mass range 8, 16 or 32 points/amu, Barchart scanning, Peak-jump collection of up to 15 peaks per scan, Complete recipe control to link multiple acquisition into a single scan, (Process Eye Professional and TOOLweb RGA only)
- Settling Time<20 ms (to 1 ppm of Maximum Signal)
- Control MethodCE operating system processor with web-server interface
- Command StructureDocumented ASCII command protocols
- Multiplier ProtectionAutomatic removal of over-range peaks from all scan modes, dedicated multiplier inhibit line for over temperature or over-pressure and x-trip filament and multiplier protect for over-pressure
- Communication10/100 Base-T Ethernet, static or automatically assigned IP addresses
- InterfacesRJ45 socket: 10/100 Base-T Ethernet, 25 way D-type female: 16 configurable TTL I/O, one dedicated multiplier over temperature inhibit input, includes power for ±15V (100 mA), +24V (100 mA) and ±3.3V (100 mA), 15 way D-type female: 4 analog in (-11 - 11 V, 22 bit), 2 analog outputs (0 - 10V, 12 bit), includes power for ±15 V (100 mA), compound output from detector pre-amplifier, 9 way D-type female dedicated vacuum gauge port: -11 - 11 V 22 bit analog input, 1 digital input (gauge OK), 1 digital output (gauge enable) and gauge power +24 V (100mA), 3.5 mm jack socket: Opto-isolated input for filament protect or control with auto detection of presence of the jack plug as failsafe, 2.5 mm jack socket: Audio output for use with speaker and wired or wireless headphones
- Number of Stored Source Settings6 sets including ion source parameters, alignment, resolution, detector calibrations, Stored in electronics and recalled by EasyView, Process Eye Professional or TOOLweb RGA
- Software Controlled Tuning ParametersElectron energy, 20 - 100 eV, Emission current, 0 - 2 mA (5 mA degas), Ion energy, 0 - 10 V, Ion extraction, 0 to -130V, Filter pole bias, 0 to -10 to +10 V
- Mounting FlangeDN 40 CF-F
- ComplianceCompliant to RoHS Directive 2002/95/EC (control unit)
Technical Specs
- TypeResidual Gas Analyzer
- Measurement TechniqueMass Spectrometry
- Mass Range100, 200, or 300 amu
- DetectorFaraday
- Faraday1.5e-11 Torr (2e-11 mbar)
- FilamentsTungsten or Thoria coated Iridium
- SoftwareBuilt-in web applications allowing RGA control and data acquisition using a platform independent web browser without the need for dedicated installed software, EasyView, Process Eye Professional and TOOLweb RGA Windows software packages for Windows 2000, XP, or Vista
- Maximum Analyzer Operating Pressure7.6e-5 Torr (1e-4 mbar)
- Ion Source Sensitivity2e-4 A/mbar
- ResolutionBetter than 10% valley for peaks of equal height across mass range
- Bake Out Temperature250°C (with extender or electronics removed)
- Operating Temperature200°C Faraday only with electronics 10 - 40°C, 150°C Multiplier with electronics 10 - 40°C and pressure <1e-8 mbar, 90°C Multiplier with electronics 10 - 40°C and pressure <1e-5 mbar
- Cross Beam Ion SourceSingle, double, or triple, depending on configuration.
- UHV Ion SourceSingle, double, or triple, depending on configuration.
- UHV Out-gassing Rate<1e-9 mbar -l/s
- PVD Ion SourceSingle, double, or triple, depending on configuration.
- Zero Blast EffectDouble & Triple Filter: <5 ppm baseline levels next to mass 2 (not applicable with 5 in. single version)
- Maximum Operating Conditions10 - 40°C, 80% RH (non condensing)
- Microchannel plate3.8e-14 Torr (5e-14 mbar)
- Single channel EM7.6e-15 Torr (1e-14 mbar) triple configuration only
- Electronics Weight on Flange1.7 kg
- Power24 VDC, 3A external universal voltage supply (included)
- Data Acquisition MethodDedicated real time acquisition processor
- Data Acquisition Speed<3 ms per point for analog scanning
- Data Collection MethodsAnalog scanning, full mass range 8, 16 or 32 points/amu, Barchart scanning, Peak-jump collection of up to 15 peaks per scan, Complete recipe control to link multiple acquisition into a single scan, (Process Eye Professional and TOOLweb RGA only)
- Settling Time<20 ms (to 1 ppm of Maximum Signal)
- Control MethodCE operating system processor with web-server interface
- Command StructureDocumented ASCII command protocols
- Multiplier ProtectionAutomatic removal of over-range peaks from all scan modes, dedicated multiplier inhibit line for over temperature or over-pressure and x-trip filament and multiplier protect for over-pressure
- Communication10/100 Base-T Ethernet, static or automatically assigned IP addresses
- InterfacesRJ45 socket: 10/100 Base-T Ethernet, 25 way D-type female: 16 configurable TTL I/O, one dedicated multiplier over temperature inhibit input, includes power for ±15V (100 mA), +24V (100 mA) and ±3.3V (100 mA), 15 way D-type female: 4 analog in (-11 - 11 V, 22 bit), 2 analog outputs (0 - 10V, 12 bit), includes power for ±15 V (100 mA), compound output from detector pre-amplifier, 9 way D-type female dedicated vacuum gauge port: -11 - 11 V 22 bit analog input, 1 digital input (gauge OK), 1 digital output (gauge enable) and gauge power +24 V (100mA), 3.5 mm jack socket: Opto-isolated input for filament protect or control with auto detection of presence of the jack plug as failsafe, 2.5 mm jack socket: Audio output for use with speaker and wired or wireless headphones
- Number of Stored Source Settings6 sets including ion source parameters, alignment, resolution, detector calibrations, Stored in electronics and recalled by EasyView, Process Eye Professional or TOOLweb RGA
- Software Controlled Tuning ParametersElectron energy, 20 - 100 eV, Emission current, 0 - 2 mA (5 mA degas), Ion energy, 0 - 10 V, Ion extraction, 0 to -130V, Filter pole bias, 0 to -10 to +10 V
- Mounting FlangeDN 40 CF-F
- ComplianceCompliant to RoHS Directive 2002/95/EC (control unit)