Precisive® 5 Application Specific Gas Analyzers
Model: PRECISIVE-5-242B
Precisive® 5 Application Specific Gas Analyzers
Model: PRECISIVE-5-242B

Overview
Overview
The Precisive® 5-242B Gas Analyzer is a real-time hydrocarbon gas analyzer based on MKS’ unique Tunable Filter Spectroscopy (TFS™) platform. Calibrated for Natural Gas composition monitoring and heating value analysis, this package is optimized for gas-phase product analysis containing C1 – C6 alkane gases as typically found in natural gas pipeline and LNG.
Technical Specs
- TypeTFS Gas Analyzer
- Measurement TechniqueTunable Filter Spectroscopy
- Measurement RangesCH4: 50-100%
C2H6: 0-20%
C3H8: 0-10%
iC4H10: 0-5%
nC4H10: 0-5%
iC5 + neoC5 (as iC5): 0-2%
n-Hexane: 0-2%
CO2:0-20%
Balance
High Heating Value (HHV)
Low Heating Value (LHV)
Specific Gravity
Wobbe Index (WI) (per ISO 6976: 2016) - RepeatabilitynC6, nC4+nC5 Channels: <0.1 mol%; All others: <0.05%
HHV, LHV & WI: <0.3% of Reading or <0.15 MJ/m3 (greater of)
Specific Gravity: <0.06% of Reading
- AccuracyGas Channels: ±0.25 mol% or ±0.5% of Full Scale (greater of)
HHV, LHV & WI: ±0.7% of Reading or ±0.25 MJ/m3 (greater of)
Specific Gravity: ±0.5% of Reading
Technical Specs
- TypeTFS Gas Analyzer
- Measurement TechniqueTunable Filter Spectroscopy
- Measurement RangesCH4: 50-100%
C2H6: 0-20%
C3H8: 0-10%
iC4H10: 0-5%
nC4H10: 0-5%
iC5 + neoC5 (as iC5): 0-2%
n-Hexane: 0-2%
CO2:0-20%
Balance
High Heating Value (HHV)
Low Heating Value (LHV)
Specific Gravity
Wobbe Index (WI) (per ISO 6976: 2016) - RepeatabilitynC6, nC4+nC5 Channels: <0.1 mol%; All others: <0.05%
HHV, LHV & WI: <0.3% of Reading or <0.15 MJ/m3 (greater of)
Specific Gravity: <0.06% of Reading
- AccuracyGas Channels: ±0.25 mol% or ±0.5% of Full Scale (greater of)
HHV, LHV & WI: ±0.7% of Reading or ±0.25 MJ/m3 (greater of)
Specific Gravity: ±0.5% of Reading