E28H 0.05-100 Torr Heated (100°C) Baratron® Capacitance Manometers with Etch Sensor
Model: e28h-heated-capacitance-manometers-etch-sensor
E28H 0.05-100 Torr Heated (100°C) Baratron® Capacitance Manometers with Etch Sensor
Model: e28h-heated-capacitance-manometers-etch-sensor
Technical Specs
Type
Vacuum/pressure Transducer
Measurement Type
Absolute
Sensor Temperature
Controlled to 100°C
Full Scale Pressure Ranges
0.05, 0.1, 0.25, 1, 2, 10, 20, and 100 Torr
Accuracy
0.25% of Reading
0.5% for <1 Torr and 0.05 Torr
Analog Output
0-10 VDC into >10K Ω load
Zero Temperature Coefficient
0.002% of Full Scale/°C for 1-100 Torr ranges
0.01% Full Scale/°C for 0.1 Torr
0.02% Full Scale/°C for 0.05 Torr
Span Temperature Coefficient
0.02% of Reading/°C
Operating Temperature Range
15-50°C
Volume
6.3 cc
Warm-up Time
2 hours for 1 Torr and higher
4 hours for <1 Torr
Overpressure Limit
45 psia (310 kPa) or 120% Full Scale, whichever is greater
Exposed Materials
Inconel®
Power Requirements
±15 VDC ±5% @ 0.6 Amps (maximum)
Compliance
CE
Technical Specs
Type
Vacuum/pressure Transducer
Measurement Type
Absolute
Sensor Temperature
Controlled to 100°C
Full Scale Pressure Ranges
0.05, 0.1, 0.25, 1, 2, 10, 20, and 100 Torr
Accuracy
0.25% of Reading
0.5% for <1 Torr and 0.05 Torr
Analog Output
0-10 VDC into >10K Ω load
Zero Temperature Coefficient
0.002% of Full Scale/°C for 1-100 Torr ranges
0.01% Full Scale/°C for 0.1 Torr
0.02% Full Scale/°C for 0.05 Torr
Span Temperature Coefficient
0.02% of Reading/°C
Operating Temperature Range
15-50°C
Volume
6.3 cc
Warm-up Time
2 hours for 1 Torr and higher
4 hours for <1 Torr
Overpressure Limit
45 psia (310 kPa) or 120% Full Scale, whichever is greater
Exposed Materials
Inconel®
Power Requirements
±15 VDC ±5% @ 0.6 Amps (maximum)
Compliance
CE
Resources & Downloads
Literature
Application Notes