Precisive® 5 Application Specific Gas Analyzers
Model: precisive-application-specific
Precisive® 5 Application Specific Gas Analyzers
Model: precisive-application-specific
Technical Specs
Type
Real-time Gas Analyzer
Measurement Technique
Tunable Filter Spectroscopy (TFS™)
Zero Drift
Not to exceed 0.2% per month; Maximum of 0.5% throughout life of bulb (~36 months)
Sample Pressure
0.1-5 psig
Flow Rate
0.1-2 L/min
Contact MKS for higher or custom ranges
Sample Cell
0.35m pathlength, 100mL volume
Weight
12kg (25.8lbs)
Power Requirements
24 VDC or 100 - 240 VAC
Operating Temperature
-20 to 50°C
-20 to 48°C (ATEX)
Communication
Modbus over Ethernet or RS485
Dimensions
Width: 17.25 in. (438.2 mm)
Depth: 6.60 in. (167.6 mm)
Height: 11.25 in. (285.8 mm)
Certifications
Class I, Division2, Group A, B, C and D, T4
Class I, Zone 2, Aex nA nC IIC T4 Gc
Ex nA nC nL IIC T4 Gc
Sira 13ATEX4316
Ex ec nC IIC T4 Gc
IECEx CSA 14.0049
Ex ec nC IIC T4 Gc
IP66, Type 4
Technical Specs
Type
Real-time Gas Analyzer
Measurement Technique
Tunable Filter Spectroscopy (TFS™)
Zero Drift
Not to exceed 0.2% per month; Maximum of 0.5% throughout life of bulb (~36 months)
Sample Pressure
0.1-5 psig
Flow Rate
0.1-2 L/min
Contact MKS for higher or custom ranges
Sample Cell
0.35m pathlength, 100mL volume
Weight
12kg (25.8lbs)
Power Requirements
24 VDC or 100 - 240 VAC
Operating Temperature
-20 to 50°C
-20 to 48°C (ATEX)
Communication
Modbus over Ethernet or RS485
Dimensions
Width: 17.25 in. (438.2 mm)
Depth: 6.60 in. (167.6 mm)
Height: 11.25 in. (285.8 mm)
Certifications
Class I, Division2, Group A, B, C and D, T4
Class I, Zone 2, Aex nA nC IIC T4 Gc
Ex nA nC nL IIC T4 Gc
Sira 13ATEX4316
Ex ec nC IIC T4 Gc
IECEx CSA 14.0049
Ex ec nC IIC T4 Gc
IP66, Type 4
Resources & Downloads
Literature
Other
Application Notes
Technical Papers
Technical Notes