300 mm Resist-Torr® XD High Sensitivity Photoresist Detection Monitor
Model: 300MM-RESIST-TORR-XD
300 mm Resist-Torr® XD High Sensitivity Photoresist Detection Monitor
Model: 300MM-RESIST-TORR-XD

Overview
Overview
The 300mm Resist-Torr® XD incorporates V-lens™ technology advancing the proven performance of the 300mm Resist-Torr® system to a new level of sensitivity, selectivity, and confidence in degas applications, compared with conventional quadrupole mass spectrometry-based residual gas analyzers. The patented V-lens™ design provides significantly improved detection power for the most challenging wafer degas applications, where low ppb detection of contaminants and photoresist is required (<15ppb specified with Argon or Nitrogen). With V-lens™ technology, the 300mm Resist-Torr XD overcomes the issues presented by challenging gases such as Argon that the limited detection power of conventional systems cannot address. In combination with the unique PR Index from MKS, this uniquely powerful platform delivers a clear advantage for the monitoring and detection of trace gases. Process engineers are able to identify costly wafer contamination earlier and maximize process yield.
Technical Specs
- Mass Range1 - 200 amu
- Ion SourceHigh sensitivity, closed ion source
- FilamentsTwin Thoria filaments
- Mass FilterDouble filter (1 in. "RF only" pre-filter with 4 in. main filter)
- DetectorDual (Faraday and Secondary Electron Multiplier)
- Minimum Detectable Concentration<100 ppb for all common gases except <10 ppm for H2
- Mass StabilityBetter than ±0.1 amu over 8 hours
- ResolutionBetter than 10% valley for peaks of equal height across mass range
- Mounting FlangeDN35CF (70 mm/2.75 in. OD) Conflat, SS capillary
- Vacuum Hardware60 l/s turbomolecular pump with high conductance analyzer housing, fast response capillary inlet system, automated vacuum controller (RVC) completely interlocked and integrated
- Analyzer Housing Base PressureBetter than 5x10-9 Torr after bakeout
- Bake Out Temperature200°C with Bakeout Jacket
- WeightApproximately 24 lbs (11.0 kg) to bolt-on process system (support stand or bracket required)
- Pneumatics60-80 psig CDA
- Electronics Weight on Flange1.7 kg
- Power Requirements88 - 264 VAC, 47/63 Hz, 600 Watts
- Maximum Operating ConditionsElectronics: 10 - 40°C, 80% RH (non condensing)
- LED Status IndicationInterlock status, filament emission, SEM, power and communications
- I/O Capability4 analog inputs and 2 outputs (plus 1 dedicated gauge input). Optional support for a large number of both analog and digital inputs and outputs, including relay control
- SoftwareProcess Eye Professional fully network compatible control platform generating under 32 bit or 64 bit Microsoft Windows XP, Vista, Server 2008 or Windows 7
- CommunicationEthernet CAT-5e
- RGA to Vacuum System Cable LengthLength 33 ft (10 m) standard RGA. Other lengths available dependent upon process system and customer requirements
- Shipping WeightApproximately 85 pounds (38.5 kg) varies with configuration
- Foreline PumpDry diaphragm standard
- TypeResidual Gas Analyzer
- InletDual path with fast-response capillary for degas monitoring and high conductance for base vacuum. Includes heater jacket.
- Other FeaturesLeak check headset socket, external filament trip socket, instrument reset
- Measurement TechniqueMass Spectrometry
- ComplianceCE
Technical Specs
- Mass Range1 - 200 amu
- Ion SourceHigh sensitivity, closed ion source
- FilamentsTwin Thoria filaments
- Mass FilterDouble filter (1 in. "RF only" pre-filter with 4 in. main filter)
- DetectorDual (Faraday and Secondary Electron Multiplier)
- Minimum Detectable Concentration<100 ppb for all common gases except <10 ppm for H2
- Mass StabilityBetter than ±0.1 amu over 8 hours
- ResolutionBetter than 10% valley for peaks of equal height across mass range
- Mounting FlangeDN35CF (70 mm/2.75 in. OD) Conflat, SS capillary
- Vacuum Hardware60 l/s turbomolecular pump with high conductance analyzer housing, fast response capillary inlet system, automated vacuum controller (RVC) completely interlocked and integrated
- Analyzer Housing Base PressureBetter than 5x10-9 Torr after bakeout
- Bake Out Temperature200°C with Bakeout Jacket
- WeightApproximately 24 lbs (11.0 kg) to bolt-on process system (support stand or bracket required)
- Pneumatics60-80 psig CDA
- Electronics Weight on Flange1.7 kg
- Power Requirements88 - 264 VAC, 47/63 Hz, 600 Watts
- Maximum Operating ConditionsElectronics: 10 - 40°C, 80% RH (non condensing)
- LED Status IndicationInterlock status, filament emission, SEM, power and communications
- I/O Capability4 analog inputs and 2 outputs (plus 1 dedicated gauge input). Optional support for a large number of both analog and digital inputs and outputs, including relay control
- SoftwareProcess Eye Professional fully network compatible control platform generating under 32 bit or 64 bit Microsoft Windows XP, Vista, Server 2008 or Windows 7
- CommunicationEthernet CAT-5e
- RGA to Vacuum System Cable LengthLength 33 ft (10 m) standard RGA. Other lengths available dependent upon process system and customer requirements
- Shipping WeightApproximately 85 pounds (38.5 kg) varies with configuration
- Foreline PumpDry diaphragm standard
- TypeResidual Gas Analyzer
- InletDual path with fast-response capillary for degas monitoring and high conductance for base vacuum. Includes heater jacket.
- Other FeaturesLeak check headset socket, external filament trip socket, instrument reset
- Measurement TechniqueMass Spectrometry
- ComplianceCE
Resources & Downloads
Drawings & CADs
300mm Resist-Torr® XD Dimensional Drawing(2.9 MB, PDF)