41C Baratron® Single-End Gauge Pressure Switches
Model: 41C13DGA2AA040
41C Baratron® Single-End Gauge Pressure Switches
Model: 41C13DGA2AA040

Overview
Overview
The 41C13DGA2AA040 Pressure Switch is ranged for 1000 Torr, has a size NW16 ISO-KF flange, 20-30 VDC Input, a 9-Pin Type "D" male subminiature electrical connector, energizes with pressure below the set point, and accuracy of 0.5% of Full scale. It is referenced to atmospheric pressure, and offers accurate and reliable protection for vacuum vacuum/pressure processes. Designed for applications where a DC signal output is not required, these switches provide relay outputs that are readily interfaced with alarms, valve actuators, computers, process controllers, load locks and other protection devices.
Technical Specs
- TypeVacuum/pressure Switch
- ConfigurationSingle-End
- Pressure Range Full Scale1000 Torr
- Fitting TypeNW16 ISO-KF
- Connector9-pin male D subminiature
- Accuracy±0.5% of Full Scale (± temperature coefficient)
- Relay ModeEnergizes above set point
- Trip Point Dead Band±3% of Full Scale
- Response Time<20 msec
- Internal Volume3.3 cc
- Temperature Coefficient±0.07% of Full Scale/°C
- Materials Exposed to Process GasesInconel and 316L stainless steel
- Overpressure2 x Full Scale or 45 psia, whichever is greater
- OutputsElectromechanical relay - SPDT (isolated) contacts rated up to 1 Amp @ 30 VDC resistive.
- Power Requirements20-30 VDC
- Ambient Operating Temperature0-50°C
- ComplianceCE Compliant to EMC Directive 2004/108/EC
Technical Specs
- TypeVacuum/pressure Switch
- ConfigurationSingle-End
- Pressure Range Full Scale1000 Torr
- Fitting TypeNW16 ISO-KF
- Connector9-pin male D subminiature
- Accuracy±0.5% of Full Scale (± temperature coefficient)
- Relay ModeEnergizes above set point
- Trip Point Dead Band±3% of Full Scale
- Response Time<20 msec
- Internal Volume3.3 cc
- Temperature Coefficient±0.07% of Full Scale/°C
- Materials Exposed to Process GasesInconel and 316L stainless steel
- Overpressure2 x Full Scale or 45 psia, whichever is greater
- OutputsElectromechanical relay - SPDT (isolated) contacts rated up to 1 Amp @ 30 VDC resistive.
- Power Requirements20-30 VDC
- Ambient Operating Temperature0-50°C
- ComplianceCE Compliant to EMC Directive 2004/108/EC