AA07B Micro-Baratron High Pressure Single-ended Pressure Transducers
Model: AA07B12PCD2GA1
AA07B Micro-Baratron High Pressure Single-ended Pressure Transducers
Model: AA07B12PCD2GA1

Overview
Overview
The AA07B12PCD2GA1 Micro-Baratron® High Pressure, Ultraclean, Single-ended Pressure Transducer offers 100 psia pressure measurement with accuracy of 1.0% of Reading and a 4 VCR-style female fitting. Electrical connections for the +13 to +32 VDC power requirements and 0-10 VDC output with a 9-pin type "D" subminiature connector at the end of a 9 inch cable. It is ideally suited for use in delivery systems that feed ultrapure gases to critical process systems. After manufacture and assembly, they are purged with ultraclean nitrogen prior to double bagging in a class 100 environment. The all-Incoloy® construction of the sensors in the AA07B allows for high overpressure tolerances that reduce errors due to line pressure spikes. High burst pressure ratings contribute to overall system safety.
Technical Specs
- TypeVacuum/pressure Transducer
- ConfigurationSingle-ended
- Measurement TypeAbsolute
- Full Scale Pressure100 psia
- Accuracy1.0% of Reading
- Analog Output0-10 VDC
- Fitting Type4 VCR female
- Electrical Connector9-pin Type D on 9 in. flying lead
- Zero Temperature Coefficient0.02% of Full Scale/°C
- Span Temperature Coefficient0.04% of Reading/°C
- Overpressure Limit150 psia
- Burst Pressure1,000 psia
- Exposed MaterialsIncoloy®
- Surface Finish≤10 µin Ra
- EnclosureStandard
- Power Requirements13-32 VDC
- Operating Temperature Range0-50°C (32-122°F)
- DisplayNo
- ComplianceCE
Technical Specs
- TypeVacuum/pressure Transducer
- ConfigurationSingle-ended
- Measurement TypeAbsolute
- Full Scale Pressure100 psia
- Accuracy1.0% of Reading
- Analog Output0-10 VDC
- Fitting Type4 VCR female
- Electrical Connector9-pin Type D on 9 in. flying lead
- Zero Temperature Coefficient0.02% of Full Scale/°C
- Span Temperature Coefficient0.04% of Reading/°C
- Overpressure Limit150 psia
- Burst Pressure1,000 psia
- Exposed MaterialsIncoloy®
- Surface Finish≤10 µin Ra
- EnclosureStandard
- Power Requirements13-32 VDC
- Operating Temperature Range0-50°C (32-122°F)
- DisplayNo
- ComplianceCE
Resources & Downloads
Drawings & CADs
AA07 & AA08 Baratron® Pressure Transducer Dimensional Drawing(1,009.5 kB, PDF)